Next Generation Visualisation & Metering Technology for Multi-phase Flows
Lead Research Organisation:
University of Huddersfield
Department Name: Sch of Computing and Engineering
Abstract
Abstracts are not currently available in GtR for all funded research. This is normally because the abstract was not required at the time of proposal submission, but may be because it included sensitive information such as personal details.
Organisations
People |
ORCID iD |
Gary Lucas (Principal Investigator) |
Publications
Laszlo Kollar (Author)
(2013)
An Analytical Method for Reconstruction of Velocity Profiles using a Multi-Electrode Electromagnetic Flow Meter
Zhichao Zhang (Author)
(2013)
Determination of power law velocity profiles by electromagnetic flow measurement
Meng Y
(2017)
Imaging water velocity and volume fraction distributions in water continuous multiphase flows using inductive flow tomography and electrical resistance tomography
in Measurement Science and Technology
Zhao YY
(2014)
Measurement and control systems for an imaging electromagnetic flow metre.
in ISA transactions
Leeungculsatien T
(2013)
Measurement of velocity profiles in multiphase flow using a multi-electrode electromagnetic flow meter
in Flow Measurement and Instrumentation
Maythem Ali (Author)
(2013)
Optimisation of an Analogue Signal Conditioning Circuit for a Novel Imaging Electromagnetic Flow meter
Kollár L
(2014)
Proposed method for reconstructing velocity profiles using a multi-electrode electromagnetic flow meter
in Measurement Science and Technology
Gary Lucas (Author)
(2013)
Two-phase flow meter for determining water and solids volumetric flow rate in inclined solids-in-water flows
Description | I. There is an urgent need for a technique for non-invasive measurement of the distribution of local axial velocity of the electrically conducting continuous phase in various types of multiphase flow. This is particularly true for performing accurate measurement of the volumetric flow rate of the water phase in 'water continuous' oil-water-gas flows, a type of flow which is frequently encountered in the oil industry. In order to measure the volumetric flow rate of the water phase, the product of the local water velocity and the local in-situ water volume fraction must be integrated in the flow cross section. It has previously been possible to measure the local water volume fraction distribution using techniques such as electrical resistance tomography (ERT) but, prior to this project, no techniques were available to measure the water velocity profile. This project has succeeded in overcoming this problem by the design of a multi-electrode electromagnetic flow meter, novel signal processing electronics and novel 'image reconstruction' methods to enable water velocity profiles to be non-invasively, accurately and rapidly measured, for what is believed to be the first time in (i) single phase flows; (ii) vertical multiphase flows (oil-in-water and gas-in-water) in which the velocity profiles are axisymmetric; and (iii) inclined three phase flows (oil and gas in water) where the velocity profiles are highly asymmetric. II. To achieve the results described above the following actions were undertaken. (i) A 50mm internal diameter electromagnetic flow meter was designed and manufactured comprising a non-conducting flow tube, an array of 16 electrodes equispaced around an internal circumference of the flow tube and a Helmholtz coil. (ii) A 'power electronic' control system was designed and constructed for supplying current to the Helmholtz coil in each of 2 modes: Mode (A) enabled a periodically reversing uniform magnetic field, orthogonal to the flow direction, to be set up in the flow cross section at the electrode array. Mode (B) enabled a periodically reversing non-uniform magnetic field, with a specifically designed magnetic flux density distribution, to be set up in the flow cross section, orthogonal to the flow direction, at the electrode array. (iii) In a parallel investigation, funded by the University of Huddersfield (UoH), a multi-channel amplifier system was designed to enable measurement of the potentials induced at the electrode array by the interaction of the flow field with the relevant magnetic field. Each amplifier circuit contained a specially designed control system which enabled the amplified flow induced potential difference, between a given electrode and a reference electrode, to be maintained within the measurement range of the analogue to digital converters, despite the presence of significant dc noise generated at the electrode/fluid interface. (iv) A variety of reconstruction techniques were designed to enable the velocity profile of the conducting continuous phase (water) to be calculated from the measured flow induced potentials at the electrode array. (A) The first approach involved setting up a system of linear equations which employed 'weight values' to relate the flow induced potential differences, between pairs of electrodes, to the water velocity in each of a series of sub-domains into which the flow cross section was divided. The water velocity in each sub-domain was then calculated from the measured potential differences using matrix inversion employing Tikhonov regularisation methods. Although successful, this approach was found to give increasingly 'noisy' reconstructed velocity profiles as the number of sub-domains was increased. (B) A far more successful approach was developed in a parallel investigation, funded by UoH, which required calculation of the Discrete Fourier Transform (DFT) of the flow induced boundary potential distribution measured at the electrode array. A mathematical technique was developed whereby the modulus and argument of each component of the complex DFT can be used to calculate the properties of a unique polynomial function relating local axial water velocity to spatial position in the flow cross section. By summing these polynomial functions, for all relevant DFT components, an analytical expression for the local water velocity distribution is obtained. Because the local water velocity is known at every point in the flow cross section, this technique represents a significant improvement over many other tomographic techniques (such as ERT) which only provide average flow properties in relatively crude sub-domains. III. The electromagnetic flow metering system described in II above was tested in single phase (water), two phase (oil-water and air-water) and three phase (oil-water-gas) flows in both vertical and inclined pipes at flow facilities at UoH and Schlumberger. Water was always the continuous phase. The system has been found to measure water velocity profiles which are in good agreement with reference measurements and with the literature. Work is ongoing to combine these measured water velocity profiles with local water volume fraction distribution profiles (obtained in the same flows by Leeds University using ERT) to enable the water volumetric flow rate Qw to be measured in axisymmetric and highly asymmetric multiphase flows . These calculated values of Qw will then be compared with reference measurements of Qw. |
Exploitation Route | 1) Use in multiphase flow meters for improving the efficiency of new and existing oil wells and in reservoir management. 2) Measurement of slurries in waste monitoring applications (e.g. in sewage processing & nuclear clean-up operations). 3) Measurement of multiphase flows in the food and chemical processing industries. 4) Use in industrial heat transfer applications in pipes, where the liquid velocity profile is combined with the temperature profile in the liquid to calculate the rate of transfer of heat energy to or from the liquid. The background intellectual property (IP) to this project which is owned by the University of Huddersfield (UoH) has already been non-exclusively licensed for use in both oilfield and non-oilfield applications to two UK SME businesses. With regard to the foreground IP generated in this project by UoH, as a result of EPSRC funding, the following applies. Under the terms of a collaboration agreement the collaborating industrial partners involved in the project are entitled to open negotiations with UoH with a view to licensing this foreground IP from UoH. The foreground IP will ultimately be used in designing key components of multiphase flow meters for use in (i) well-head and down-hole monitoring of multiphase oil wells; and (ii) measuring multiphase flows in other process industries such as the nuclear industry and the food processing, chemical processing and mineral processing industries. |
Sectors | Chemicals Energy |
Title | MEANS AND METHOD FOR MONITORING THE FLOW OF FLUID |
Description | The invention relates to an means and method for monitoring the flow of a fluid. The invention relates to an electromagnetic flow meter and method for measuring the axial velocity profile of a conducting fluid. The conducting fluid may be a conducting single phase fluid or a conducting continuous phase of a multiphase fluid. The conducting fluid may have a uniform flow profile or a non-uniform flow profile. The electromagnetic flow meter and method measure the axial velocity profile of a conducting fluid by dividing the flow cross section into multiple pixels and determining the axial velocity of the conducting fluid in each pixel. Having. derived the axial velocity profile, the electromagnetic flow meter and method may further derive the volumetric flow rate of the conducting fluid. The electromagnetic flow meter and method may be suitable for measuring the axial velocity profile and optionally the volumetric flow rates of each phase of a multiphase fluid. |
IP Reference | WO2011128656 |
Protection | Patent application published |
Year Protection Granted | 2011 |
Licensed | Yes |
Impact | Further research funding to develop technique for medical imaging. |
Title | Means and Method for Monitoring the Flow of Fluid |
Description | See also WO2011128656 |
IP Reference | US20130036817 |
Protection | Patent application published |
Year Protection Granted | |
Licensed | Yes |
Impact | Further research funding has been obtained to apply the technique to medical imaging |